Since temperature affects the measuring accuracy of a pressure sensor, there always remains a little temperature error in the rated temperature range despite a variety of compensation measures. เกจวัดแรงดัน4บาร์ is frequently expressed in the data sheets of manufacturers of pressure sensors as temperature coefficient (abbr. TC). This coefficient describes a (linear) error, beginning with a reference point, which generally is room temperature.
Accordingly, the temperature error at room temperature is zero and increases with increasing difference of the temperature from room temperature with the specified coefficient in linear fashion (see figure). Basically, not merely the zero-point temperature coefficient (abbr. TC0), but also that of the span (TCC) must be taken into account individually. The zero-point error describes the effect of temperature on the zero signal. The error of the span specifies the effect of temperature on the full scale value. The average person temperature coefficients of zero point and of span are often specified as magnitudes, meaning that they are often either positive or negative.
If within an individual instrument the zero-point error gets the same sign because the error of the span, both of these temperature errors may accumulate in worst case. A typical value for the average zero-point temperature coefficient of a pressure sensor is: 0.2 % / 10 K. Link to WIKA pressure sensors.